Semi Preparative Binary High Performance Liquid Chromatography HPLC Instrumentation
Description:
LC3000 I semi-preparative gradient system is designed for laboratory processing flow gradient elution conditions 50mL/min purification. The flow range suitable for inner diameter is 20 to 30mm chromatographic column, which can meet the single purified milligram to gram purification demand samples obtained, the purity and yield of the best. It is mainly used for separation and purification to gradient elution sample preparation. LC3000 I semi-preparative gradient system is also used in the preparation method validation and optimization. It will analyze the resulting liquid chromatography preparative chromatography conditions magnified to semi-preparative chromatography system; verify preparative chromatography conditions to further optimize the recommendation system.
Ordering Details:
Item |
PN |
Product description |
QTY |
Unit |
LC 3050 semi-prep HPG HPLC SYSTEM |
1 |
set |
||
1 |
L.F.P005 |
Semi-Preparative HPLC Pump ,with hump head 50ml |
2 |
pcs |
2 |
L.F.UV3023 |
UV 3000 UV Detector(Semi-Prep),with 50ml flow cell |
1 |
pcs |
3 |
L.F.D001 |
Dynamic Mixing Chamber,Preparative |
1 |
pcs |
4 |
I.K.A.A1359 |
A1359 Manual injection valve, 6-port/3-channel, stainless steel, 1/8", 300 bar, incl. accessories,Semi-Preparative |
1 |
pcs |
5 |
A.G.103 |
CXTH 3000 Chromatography Workstation |
1 |
pcs |
6 |
L.F.S.092 |
Start-up Kit for Binary Semi-Preparative |
1 |
pcs |
7 |
L.F.R3000 |
Solvent tray,analytical |
1 |
pcs |
Note:
* When the customers use the HPLC system for the first time, we suggest that solvent filter,
vacuum pump, ultrasonic cleaner, AC regulated power supply should be equipped.
Applications:
Purify products from micro gram level to gram level
Explore purification method for large scale purification
Technical parameter:
P3000 High-Pressure Infusion Pump
Technical Parameters
Flow Rate Range: ( 0 - 49.99 ) mL/min
Max Working Pressure: 25 MPa
Flow Accuracy: ± 2 % ( 5 mL/min)
Flow Repeatability: RSD ≤ 0.1 % ( 1 mL/min )
PSU: 220 V ± 22 V, 50 Hz, 120 W
Dimensions and Weights: 450 × 270 × 180 mm( L × W × H ) 12.8 kg
Wavelength Range: ( 190 - 700 ) nm
Light Source: D2 lamp (standard configuration), Tungsten lamp (optional available)
Band Width: 8 nm, Wavelength accuracy: ± 2nm, Wavelength repeatability:0.4nm
Baseline Noise: 1.5 × 10-5 AU/h
Baseline Drift: 4×10-4 AU/h
Dynamic: Velocity of flow 1mL/min; Methanol, wavelength 254 nm
Detection Cell Volume: 7 uL
PSU: 220 V ± 22 V, 50 Hz, 150 W
Dimensions and Weights: 450 × 270 × 180 mm ( L × W × H ) 10.4 kg
Competitive Advantage: